29
SillHa Oral Wellness Instrument
Apply one drop of sample to the center of
each pad starting with the yellow pad and
proceeding to the other end.
Start timing. Tap [Start] immediately after
the dropping of the sample is complete
(within 5 seconds) (a). Then, open the
reading unit cover (b).
• The instrument starts counting down
from 10.
REFERENCE
To stop measurement: Tap and hold [Stop].
During the 10-second countdown (before
the beep stops sounding), tilt and blot one
edge of the test strip on a clean absorbent
surface to remove excess sample from all
pads.
REFERENCE
When three seconds are left until the end of the countdown, a short beep starts
sounding, and a long beep sounds at 0 second.
IMPORTANT:
•
Finish dropping the sample on all eleven
pads within 10 seconds. If it is delayed,
results may not be accurate.
•
Apply one drop of sample, keeping the tip
of the dropper away from the pad. If the
dropped volume is too much or too little
for measurement, results may not be
accurate.
•
Perform the operation in next step after
completing the dropping of the sample.
If there is a delay in starting
measurement, results may not be
accurate.
Reading
unit cover
(a)
(b)
2.3.3
Summary of Contents for SillHa LH-4912