SE-1520 ■ 3-13
3-3
パラメータメニュー
Chapter 3 Sub Menu
3-3-1
パラメータ設定内容の印字
Normal value ranges
4.
Enter the lower value of the normal
value range by using the numeric
keys.
Press [ENTER] key and the cursor
moves to the entry position for the
highest limit.
Enter the upper value using the
numeric keys and press [ENTER] key.
Then the next setting screen is
displayed. After the normal value
ranges for blood and urine are
entered, Correction Temperature
Setting Screen is displayed.
●
●
●
Set coefficient factors
3.
Enter the coefficient factor "A"
by using the numeric keys.
Any number from 0 to 10000
can be entered.
P r e s s [ E N T E R ] k e y a n d t h e
c u r s o r m o v e s t o t h e e n t r y
position for "B".
Enter a coefficient factor "B" and
press [ENTER].
Any number from -10000 to
10000 can be entered. Press
[ E N T E R ] k e y a n d t h e n e x t
setting screen is displayed.
After coefficients for whole
blood, serum, plasma and urine
are entered, the Normal Value
R a n g e S e t t i n g s c r e e n i s
displayed.
●
●
Enter minus signs and
decimal points by using the "-
/." key.
To enter a minus sign, press
[-/.] before pressing any
numeric keys. A decimal
point can be entered. Press
[-/.] key after any numeric key
is pressed.
e.g.) To enter "-12
Press [-/.] [1] [2]
e.g.) To enter "3.4"
Press [3] [-/.] [4]
e.g.) To enter ".5"
Press [0] [-/.] [5]. If only [-/.]
[5] is pressed, -5 is entered.
If the wrong key is pressed,
press [MENU] key and [-/.]
simultaneously to delete the
last entered character.
If [START] key is pressed,
the initially displayed value
is restored.
▲
B l o o d N ORM L〈 1 . 0 0 0 〉
S - 0 2
U〈 1 5 . 0 0 〉
S a v e ?
Y e s( S T A R T ) N o( S T O P )
▲▲
S e r u m
a〈
1 . 0 0 0 〉
S - 0 2
b〈
0 . 0 0 0 〉
S e r u m
a〈
1 . 0 0 0 〉
S - 0 2
b〈
0 . 0 0 0 〉
Set the correction temperature
5.
P r e s s [ - ] k e y t o s e l e c t
Correction Temperature.
Press [ENTER] key. The Entry
Check Screen is displayed.
●
●
T e m p e r a t u r e
[ −−−− ]
S - 0 2
B l o o d N ORM L〈 1 . 0 0 0 〉
S - 0 2
U〈 1 5 . 0 0 〉
Only the enzyme items can
be converted at the selected
correction
temperature.
▲