2-6
Measurement Result
Chapter 2 Measurement
2-6-1
Printing the Normal Measurement Results
(1)Normal printing
1
2
6
3
4
5
7
Measurement date
The format of the measurement date is date format .
ID
When there is no ID, it becomes blank. ( ID:
is not printed.)
Concentration value
When the measurement result is out of the
measurement range, the following is printed.
When the measurement result is over the range: the upper limit of the
measurement range are printed.
When the measurement result is under the range: the lower limit of the
measurement range are printed.
Error mark
Stability error "?"
Error of the lower limit of the measurement range "<"
Error of the upper limit of the measurement range ">"
Error of the lower limit of the normal measurement range "▼"
Error of the upper limit of the normal measurement range "▲".
* All fields for concentration values are printed with "-" in the event of no
solution errors.
Sample type
Printed in accordance with the selected language.
Whole blood measurement: Whole blood
Plasma measurement: Plasma
Serum measurement: Serum
Urine measurement: Urine
The lot information for used E-Plate is printed out.
Error printing
When error message is printed at the end of measurement
result, see 5-1 Error Messages .