Calibration
Chapter 2 Measurement
Calibration is necessary for maintaining measurement
accuracy. By performing calibration, the problems caused by
differences among lots are solved and measurement accuracy
is maintained.
Basically, calibration is performed for each exclusive plate
type using exclusive calibrators. However, it will take a long
time to perform calibration for all exclusive plates.
To reduce the time and trouble, this system performs
c a l i b r a t i o n b y r e a d i n g m a g n e t i c c a r d s r e c o r d i n g l o t
information of each exclusive plate.
This method is called Magnetic Card Calibration.
Using this magnetic card calibration method, calibration is
performed by inserting the magnetic card attached to the
exclusive plate into the magnetic card reader.
2-5-2
Magnetic Card Calibration
D i f f e r e n c e s a m o n g l o t s o f t h e e x c l u s i v e p l a t e a r e
automatically corrected by inserting the magnetic card
attached to the exclusive plate into the magnetic card reader.
If the measurement is performed when the magnetic card calibration is not
performed correctly, errors occur and measurement may not start.
In this case, be sure to perform
Magnetic Card Calibration.
Check the recorded information
1.
Press [3] on the Main Menu.
T h e C a l i b r a t i o n M e n u i s
displayed.
Press [2].
Lot information of all items
r e c o r d e d i n t h e s y s t e m i s
printed.
When the printing is completed,
you return to the Calibration
Menu.
●
●
1 . C a r d
2 . I n f o
( 1/1 )
P r i n t i n g
…
SE-1520 ■ 2-23