b
2-20 ■ SE-1520
Set the sample type (when necessary)
6.
S e e p r o c e d u r e 6 o f 2 - 3 - 3
Checks before Measurement
to set the sample type.
●
2-4
Measurement
Chapter 2 Measurement
The entered ID is displayed.
Press [ENTER].
The entered ID is stored and
t h e s c r e e n r e t u r n s t o t h e
Ready Screen.
●
●
●
When the barcode reader is
used, a message
Barcode
r e a d e r i s a v a i l a b l e
i s
displayed.
▲
Set the E-Plate on the table
7.
Set the E-Plate onto the proper
location of the analyzer.
●
Make sure the lot number
8.
When E-Plate is set, either
following 【1】 or 【2】 is
displayed.
When this screen is displayed,
check if used lot number for E-
Plate is same as △△△△△ on
the display.
●
●
A p p l y s a m p l e
■■■
[ W h o l e B .
]
△△△△△
If the lot number is not same, a correct data can not be obtained.
In this case, calibrate by a magnetic card.
When this screen is displayed,
p r o c e e d t o t h e n e x t s t a g e
without checking lot number.
●
A p p l y s a m p l e
■■■
[ W h o l e B .
]
N o . ****
I D<*****
■■■■
>
O K( E N T E R )
If not enter measurement No. and ID, to 7. Set the E-Plate on the table .
Normally measurement No. is allocated automatically. ID is not entered in the such way.
【1】
【2】