MAGB2 User Guide
5
V1.7 2021-10-01
2. Installation
2.1. Remote or Compact?
Any MAGB2 flowmeter can be delivered in two versions; Compact or Remote. The compact version has the
transmitter unit connected directly to the sensor body. This version does not require any further mounting or
installation of the transmitter.
The remote version has a separated transmitter. It is connected to the sensor via a cable. The cable entry
into the sensor is protected by a junction box, which is potted to IP68. The cable entry on the transmitter
side is through a M20x1,5 gland.
The cable type used for the connection between sensor and transmitter for remote versions:
UNITRONIC® Li2YCY (PiMF), 2x2x0.34
The MAGB2 can be used with a maximum 6 mts of cable between sensor and transmitter in remote version.
2.2. Sensor installation
Sensor dimensions can be found on chapter 17.
Proper sensor installation is extremely important in order for your flowmeter to work correctly. Below, you
will find the minimum sensor installation requirements that need to be respected at all time.
Sensor
installation
requirements: