11. Electromagnetic Compatibility (EMC)
Caution
The use of non-approved accessories may result in increased emissions
or reduced immunity of the equipment. A list of approved accessories is
provided in Chapter 5 of this manual.
When this equipment is used adjacent to other electronic devices, the
user should observe the equipment to verify normal operation.
Guidance and manufacturer’s declaration – electromagnetic emissions
The bed is intended for use in the electromagnetic environment specified below. The customer or the
user of the bed should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The bed uses RF energy only for its internal function.
Therefore its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The bed is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
746-574-AM-1
59