Electrical data
Power input
1.5A max. 120V a.c. 60Hz
Duty rating
Intermittent 10% (2 min. on, 18 min. off)
Safety standards USA/Canada
CERTIFIED AS TO ELECTRICAL SHOCK,
FIRE AND MECHANICAL HAZARDS ONLY
IN ACCORDANE WITH:
CAN/CSA-C22.2 No. 60601-1 (2008)
CAN/CSA-C22.2 No. 60601-2-52 (2011)
ANSI/AAMI ES60601-1 (2005)
IEC 60601-2-52 (2009)
Electric shock protection
Class II
Type B
EMC
Complies with EN 60601-1-2
Backup battery
24V, 1.2Ah sealed rechargeable lead/acid gel
Typical lifetime
4 years max.
Backup duration
6 full lift / lower cycles with SWL
Liquid ingress protection
IP24 (2-Protection against solid particle
ingress larger than 12.5mm – fingers or
similar objects. 4-Protection against liquid
ingress – water splashing against the
enclosure from any direction shall have no
harmful effect).
Environmental protection
This product is marked with the crossed out wheelie bin symbol to
indicate that it is covered by EU Directive 2002/96/EC on Waste
Electrical and Electronic Equipment (WEEE). This symbol indicates
that all electrical and electronic products must be taken to a dedicated
collection point at end of their working life. Contact ArjoHuntleigh or
your local authority for advice on proper disposal.
Transportation and storage
Handle with care. Do not drop. Avoid shock or violent impact.
This equipment should be stored in a clean, dry and well-ventilated area which
meets the following conditions:
Temperature
-10
°
C to 50
°
C (-14
°
F to 122
°
F)
Relative humidity
20% to 90% non-condensing
Air pressure
800 to 1060hPa
Caution
If the bed is stored for a long time, every three months it should be
connected to the electricity supply for 24 hours to recharge the backup
battery. Failure to do this may reduce the life of the battery.
746-574-AM-1
57