Service Procedure 21 -
Replacing the Hand Control
1)
Remove the battery pack.
2)
Unplug the hand control from the
socket located on the left side of the
control box.
Fig. 37
3)
Remove the top screw of the handle
with a 4mm Allen key and slide out the
wire from the bracket (see Fig. 38).
Fig. 38
4)
Install the new hand control and attach
the cable to the hand control clip (see
Fig. 38).
5)
Replace the battery pack.
6)
Test if the hand control is working
properly as per service procedure 4.
Hand Control socket