Service Procedure 1 - Check for Updates
Service Procedure 2 - Checking the Battery and Charger
Checking the Battery
ArjoHuntleigh has added a fuse in battery NDA0100-20.
There is no noticible difference between the former and the new battery version other than the labeling
(see following pictures for more details on the former battery labeling and the new battery labeling).
Only battery testers with current less than 20A, such as battery tester part #8662950 (refer to TAN
SEE201204 and see Fig. 9 on next page), can be used for the battery that includes the fuse to prevent
this component from blowing and making battery unusable.
Battery tester #8662950 can also be used on the battery without fuse.
The battery tester will load the battery with a high current for a short time. This will show the battery
•
Check if there are any field correction bulletins, safety
notices or technical bulletins that have been published
since the last service was performed.
•
This verification must be done to keep the product up to
date according to safety and product improvements.
Bulletins and notices can be generated as a result of an
engineering change note, a safety incident report, a change
to form/fit, etc.
Fig. 7
Fig. 8
Former labeling
(battery without fuse)
New labeling
(battery with fuse)