Service Procedure 15 - Replacing the Battery Discharge
Indicator PCB
7)
Put back battery and test battery discharge indicator to confirm that it works properly.
1)
Remove control box cover.
2)
Unplug J5 and J7 connectors as shown on electric
diagram from battery discharge indicator PCB.
3)
Unscrew both nuts holding the battery discharge
indicator PCB using a 5mm socket, and remove
device.
4)
Put a new battery discharge indicator PCB in place
as shown in Fig. 30, fastening the two nuts and flat
washers securely. Do not over-tighten the nuts
holding the circuit; as too much torque could
damage the board. Screw in the cover.
5)
Plug J5 and J7 connectors as shown on electric
diagram on battery discharge indicator PCB.
6)
Install control box cover.
Fig. 30
Warning: If any component of the electrical system is disassembled, it is essential that all parts
are reinstalled correctly. If not, this could result in battery heating which could lead to, hazardous
product leakage, fire and explosion. Exposure to those risks could lead to death or serious injury.
J5
J7