61
Citadel Bed Frame System
Fault Indications
The bed’s control software indicates problems in the electrical system by means
of flashing indicators on the Attendant Control Panel (ACP). If you experience
any of the indications below, contact ArjoHuntleigh or an ArjoHuntleigh-approved
service agent.
Indication
Possible cause
ACP deck height and head down tilt indicator LEDs flashing
Height actuator fault (head end)
ACP deck height and foot down tilt indicator LEDs flashing
Height actuator fault (foot end)
ACP backrest indicator LED flashing
Backrest actuator fault
ACP thigh section indicator LED flashing
Thigh section actuator fault
ACP calf section indicator LED flashing
Calf section actuator fault
Deck height, head down tilt, backrest and thigh section indicator LEDs
flashing
Control unit fault
Product Lifetime
The lifetime of this equipment is typically ten (10) years. Lifetime is defined as
the period during which the product will maintain the specified performance and
safety, provided it has been maintained and operated in conditions of normal use
in accordance with the requirements in these Instructions for Use
Accessories
Recommended accessories for the
Citadel
range are shown in the table below.
Note that some items may not be available in all countries.
Accessory
Product code
Lifting pole with strap and handle
ENT-ACC01
IV pole
ENT-ACC02
Angled IV pole
ENT-ACC04
Fracture frame
ENT-ACC05
Syringe pump holder
ENT-ACC07
Oxygen bottle holder (PD cylinder)
ENT-ACC08
Small traction assembly
ENT-ACC10
IV bag holder for lifting pole
ENT-ACC13
Additional 2 hooks for IV pole
ENT-ACC14
Oxygen bottle holder (B5 cylinder)
ENT-ACC18
Urine bottle holder
ENT-ACC19
Heavy duty IV pole
ENT-ACC24
Transducer mounting pole
ENT-ACC26
Head end traction assembly
ENT-ACC32
Head end ITU panel
ENT-ACC34
Monitor shelf
ENT-ACC64
Mattress storage bag
ENT-ACC80
Hanger bracket
ENT-ACC81