57
Citadel Bed Frame System
PREVENTIVE MAINTENANCE
This product is subject to wear and tear during use. To make sure that it
continues to perform within its original specification, preventive maintenance
procedures should be carried out at the intervals shown.
This list indicates the minimum recommended level of preventive
maintenance. More frequent inspections should be carried
out when the product is subjected to heavy use, aggressive
environments or where required by local regulations.
Failure to carry out these checks, or continuing to use the
product if a fault is found, may compromise the safety of both
the patient and caregiver. Preventive maintenance can help to
prevent accidents.
Actions to be completed by caregiver
Daily
Weekly
Check operation of side rails
Visually check casters
Check operation of the CPR backrest release handles on both
sides of the bed
Visually check power supply cord and mains plug
Carry out a full test of all electrical bed positioning functions
(backrest, height, tilt, etc.)
Check that the patient controls, caregiver controls and Attendant
Control Panels operate correctly
Check that the weighing system controls operate correctly
Check operation of the anti-entrapment system and clean the
sensor lenses (see page 44)
Check the mattress for damage and fluid ingress
Examine the lifting pole, strap and handle
If the result of any of these tests is unsatisfactory, contact ArjoHuntleigh or an
ArjoHuntleigh-approved service agent.
The procedures in the following table must be carried out by
suitably trained and qualified personnel. Failure to do so may
result in injury or an unsafe product.