42
Citadel Bed Frame System
VariZone Patient Movement / Exit Detection
The patient movement detection function should be checked
periodically for correct operation and before each new patient
uses the bed.
Mattresses that are not approved by ArjoHuntleigh should be
validated by the user to make sure correct operation with the
VariZone system.
The patient movement detection system can be set to alarm when undesired
movement of the patient occurs. The sensitivity of the patient movement
detection, relative to the center of the deck, can be varied incrementally.
The controls for the patient movement detection system are located on the foot
end split side rails.
Figure 29: Patient movement detection controls
In Bed
- this button activates / deactivates patient movement detection
and increases the sensitivity of the system.
Patient Movement Detection Threshold Display
- an
indicator shows the current system status and the selected
sensitivity of patient movement detection.
Egress
- this button activates / deactivates patient movement
detection and decreases the sensitivity of the system.
Adjusting the deck profile while patient movement detection
is active can trigger an alarm if patient movement detection
is set to a high sensitivity.
VariZone Mute Button
- When the
VariZone
alarm is triggered,
pressing the mute button will silence the audio portion of the alarm for
three minutes. When the
VariZone
alarm is triggered and is muted,
adding the patient /- 10% back onto the bed will clear the
alarm and re-enable
VariZone
to the last
VariZone
setting.
.
.