SH-7000-Chroma-EN-04/17
34
6
Initialisation and Updates
Because each pump works with two processors, it has to be ensured that both have the
same SW installed. Spare parts PCB are sent with the newest SW release and need to be
updated, initialized and calibrated by the service personnel.
6.1
Release Note
PremiumLink™ is arcomed ag own pump update software tool. This program undergoes
constant improvements and has different releases. New releases are announced with
arcomed ag’s TecUpdate bulletin or can be requested at [email protected]
!!WARNING!!
Due to the risk of failure during updates, only trained and qualified personnel (certification)
may carry out these procedures. Any damage done onto the pump due to false handling
do not fall under warranty terms. There are different technical trainings for the use and
handling of P
remiumLink™. Do
not
use the following instructions for the
Premium
generation pumps. The instructions
apply only to Chroma
generation devices.
NOTE:
Following instructions describe the updates using the standard PremiumLink™ version. A
standard sw update
and the update procedure after exchanging PCB’s are described.
Summary of Contents for Volumed nVP7000 Chroma