Trouble shooting
Fault
Cause
Action
Over Infusion
Incorrect giving set
Use Arcomed approved set
Speed tuning requires adjustment Adjust Speed tuning (address 10)
Wrong set selection (in VP7000
silicone is PVC selected)
Select correct IV brand (address
30)
Worn peristaltic cover
Replace cover
Worn door hinge
Overhaul door
Under infusion
Incorrect giving set
Use Arcomed approved set
Speed tuning requires adjustment Adjust Speed tuning (address 10)
Wrong set selection (in VP7000
PVC is silicone selected)
Select correct IV brand (address
30)
Worn out peristaltic cover
Replace cover
Worn out door hinge
Overhaul door
High Occlusion
Incorrect giving set
Use Arcomed approved set
Pressure gain (address 15)
adjustment required
Adjust Pressure gain
Pressure Transducer requires
mechanical adjustment
Adjust transducer
Pressure Transducer damaged
Replace transducer
Low Occlusion
Incorrect giving set
Use Arcomed approved set
Pressure gain (address 15)
adjustment required
Adjust Pressure gain
Worn door hinge
Overhaul door
Pressure Transducer requires
mechanical adjustment
Pressure Transducer requires
mechanical adjustment
Pressure Transducer damaged
Replace transducer
Will not power
up (on mains)
Blown fuses
Check and replace
Wire detached from mains socket Reconnect
Display cable defect
Check and replace Display PCB
Transformer faulty
Check and replace main PCB
Fuse F1 defect
Replace Main PCB and send it to
Arcomed with comment that
probably F1 could be broken.
Warning:
If F1 will be manually
replaced with a soldering gun,
warranty will be not given.