Address
Description
Default
18
Stop Flow
14
– 18:
Internal Stop Flow
114
– 118:
Robson Stop Flow
15
19
Motor torque. Internal setting
230
2.4
Optional Settings
Address
Description
Default
20
Backlight on mains, 4099 = 40% night 21h-7h, 99% day
4099
21
Backlight on battery, 4070 = 40% night 21h-7h, 70% day
4070
22
Near empty alarm in seconds. Triggers pre-alarm VTBI is
reached. Up to 180 seconds.
0
23
Stand by delay in seconds. If the pump is left stopped, a
reminder alarm will be triggered after the stand by delay.
240
24
Acoustic alarm, ex: 3569: 3: pitch height of the sound, 5 *
100 ms interval between alarm beeps, 6: acoustic volume
for key beep, 9: acoustic volume for alarms.
3569
25
Preset Bolus rate in ml/h, internally limited to: PVC 999.9
ml/h, Silicone: 1200 ml/h
1200
26
Max. Bolus volume in ml. The volume for an automatic bolus
can’t exceed this value.
10
27
Max. Rate in ml/h. Absolute max rate of the pump including
rate during bolus applications. Internally limited to: PVC
999.9 ml/h, Silicone: 1200 ml/h
1200
28
KVO rate in 0.01 ml/h, max. value 10ml/h
300
30
0: IV-Set PVC
1: IV-Set Silicon
2: Low Volume (rate limited to 700 ml/h, Adr. 10 adjustable
to 750)
3: IV-Set PVC low volume. Compensates for low volume
sets, redo volume test after change.
*
34
Language
0: first language, 1: second language
0
35
PCA Password (access to PCA parameters)
1001
36
Pressure limit (default pressure limit)
600
37
Max. Pressure limit (maximum pressure limit)
1000
40
Air limit in µl to trigger an air alarm (50 to 250µl)
150