![Aquaread Aquaprobe AP-5000 Instruction Manual Download Page 55](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-5000/aquaprobe-ap-5000_instruction-manual_2960643055.webp)
Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
Gently pour 300mL of the solution into a clean calibration tube and drop the Probe in all the
way. Again, agitate and swirl the Probe and bottle several times in order to remove any air
bubbles that may be clinging to the Turbidity electrode.
Follow the procedure detailed above for Zero point calibration as far as step 6, then select
either 20 or 1000, dependant upon the solution the probe is in. Wait while the Meter
stabilises and calibrates.
After successful calibration, the ‘Calibrating 100%’ screen will be displayed along with the
Calibration Report, which will show the voltage output from the Turbidity Receiver Electrode
in millivolts (mV). Press the
OK
key to continue.
Rinse the probe thoroughly then repeat this procedure for the third point.
15.4.9.
Errors During Calibration
If a problem occurs during calibration, an error message will be displayed. Refer to
in section 10 for error message handling.
15.4.10.
Lens and Sleeve Maintenance
On a daily basis, the lenses on the electrode should be wiped over with a soft damp cloth.
Similarly, the inside of the Probe Sleeve and Sleeve Cap should be kept clean and free
from any deposits that may cause stray reflections.
Never use an abrasive cleaner on the inside of the Probe sleeve or cap as they have
been treated with a non-reflective coating which can be easily damaged.
The inside of
the sleeve should be wiped over with a soft damp cloth and
non-abrasive
detergent.
Always re-calibrate the zero point after cleaning the sleeve or lenses.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 55 of 138