Important
If you have ISE electrodes fitted to your Aquaprobe
®
, you must now remove the rubber
sealing caps from all the ISE electrodes.
Now the DO 100% saturation point should be checked and if necessary, calibrated in damp
air.
To Check / Calibrate the 100% Saturation Point in Damp Air
1. After calibrating with RapidCaL, remove the Probe from the cup, wash in fresh
water, then shake off ensuring there are no droplets adhering to the DO membrane.
2. Moisten a clean cloth or piece of tissue paper with fresh water and wrap it around
the end of the probe ensuring all the holes are covered. Place the probe on a flat
surface. Do not hold the probe, the heat from your hands will warm the probe up and
interfere with calibration.
3. Wait until the temperature and DO
measurements are
completely stable
.
This is
very
important
.
If the DO measurement is 100%
+/- 1%, there is no need to recalibrate.
4. If recalibration is needed, refer back to the
screen shown in item 5 above and select DO
100%.
5. Wait while the Aquameter
®
carries out the
calibration procedure.
6. When the ‘Calibrating 100%’ screen (shown
above) is displayed, press OK then ESC repeatedly to return to normal reading
mode.
Finally, if you have any optical electrodes fitted, you should calibrate the zero points now
using fresh, still mineral water. Refer to the relevant sections in this manual for advice on
zero point calibration of your optical electrodes.