27.3. AP-5000 Installation into the Flowcell
Grease the threaded section and O ring of the Adapter
Ring and screw the assembly into the Flowcell.
27.4. Flowcell Operation
Connect the Flowcell to a pumping device so that
sample water enters at the bottom and exits at the top.
Adjust the flow rate so that there is no visible
turbulence or cavitation within the tubes.
Connect an Aquameter and monitor the readings. If the
readings are jumpy or erratic, reduce the flow rate. The
ideal flow rate is around 30 litres/hour (8 US
gallons/hour), although the Aquaprobe is capable of
operating at flow rates as low as 15 litres/hour (4 US
gallons/hour). Flow rates above 60 litres/hour (16 US
gallons/hour) are not recommended.
27.5. Caution
The maximum recommended operating pressure of the
Flowcell is 500mb (7.25 PSI). Select your pumping
device accordingly. If necessary, use a three-way
bypass valve so that this limit is not exceeded. Tip: use
the depth measurement feature of the AP-5000 to
measure the operating pressure. At 500mb, the AP-
5000 will show a depth of approximately 5 meters.
27.6. Flowcell Cleaning
After use, rinse the Flowcell thoroughly with fresh
water. To remove stubborn deposits, scrub the inside of the Flowcell with a bottlebrush and
non-abrasive detergent, then rinse thoroughly.
Never clean the Flowcell with solvents, alcohol or concentrated acid/alkaline based
cleaning products such as Decon 90. These products can strip the anodised finish
from the Flowcell and damage the plastic and rubber components.
27.7. Flowcell Troubleshooting
Problem
Cause / Remedy
DO readings are abnormally
high or are fluctuating wildly.
✔
Aeration of sample water. Check all joints for air
leaks. Reduce flow rate to avoid cavitation.
Turbidity or other optical
electrode readings are
abnormally high or are
fluctuating wildly.
✔
Air bubbles adhering to the Turbidity Electrode
lenses. Agitate Flowcell to dislodge.
✔
Aeration of sample water. Check all joints for air
leaks. Reduce flow rate to avoid cavitation.
✔
Optical electrode not zeroed after adapter ring
fitted. Zero the optical electrode.