1. Pour 300mL of RapidCal solution into one of
the calibration cups provided as shown in the
adjacent photograph.
2. Remove the storage cap from the pH
electrode if fitted, wash the Probe in distilled
water, then gently lower the Probe into the
calibration cup.
3. Bang the probe against the bottom of the cup
several times to dislodge any air bubbles.
4. Wait until all readings are
completely
stable.
The longer you can leave the probe to achieve
thermal equilibrium before proceeding, the
better. A minimum of two minutes is
recommended.
5. Ensure the temperature of the solution is
between 5ºC and 40ºC (41ºF – 104ºF). The
closer to 25ºC the better.
6. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
7. Select
RapidCal.
The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until all readings are stable, then it will send the RapidCal command to
the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Calibrating
100%
Press [OK]
When calibration is complete, press
OK
then
ESC
to return to normal reading mode
.