15.4.6.
Turbidity Zero Point Calibration
To calibrate the Turbidity zero point (zero the electrode), follow these steps:
1. Put at lease 300mL of clean water in a calibration tube (bottled still mineral water is
recommended), remove the storage cap from the pH electrode if fitted, wash the
Probe in clean water, then drop the Probe in all the way.
The Sleeve End Cap must
be fitted
.
Agitate the Probe and swirl the bottle several times in order to remove any air
bubbles that may be clinging to the Turbidity electrode.
2. Switch the Aquameter
®
on and wait until the temperature and turbidity readings are
stable. If the turbidity reading is very high, there are probably air bubbles adhering to
the lenses. Agitate the Probe to remove.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6.
Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:TURB
l
4:EMPTY
2:EMPTY
l
5:N/A
3:EMPTY
l
6:N/A
The TURB electrode should have been assigned to AUX socket 1 when it was fitted. Press
the OK or right arrow key to select TURB. The screen will change to:
CALIBRATE TURB
ZERO? [01/Jan/14]
1000? [01/Jan/14]
20? [01/Jan/14]
The dates shown to the right of each point are the dates of the last successful calibration.