13. pH/ORP Electrode Calibration and Maintenance
13.1. Recognising the pH/ORP Electrode
The combined pH/ORP electrode is easy to recognise because it is the only electrode that
is not black. This electrode has a clear, gel filled body.
13.2. Electrode Removal and Replacement
The pH/ORP electrode can be unscrewed from the Probe body by rotating it anti-clockwise.
When replacing an electrode, apply a little silicone grease or similar lubricant to the thread
and O ring, then screw fully in.
Gripping the knurled collar at the top of the electrode, tighten until the O ring is fully
compressed.
Useful Tip:
The red lanyard that is attached to the pH/ORP storage cap makes a very
useful belt wrench for tightening and loosening the pH/ORP and AUX electrodes.
Slide the lanyard over the electrode and use it to grip the knurled body.
Never immerse an AP-5000 with the pH/ORP electrode removed. This will cause serious
damage to the electrode socket.
This is not covered by your warranty.
13.3. Keeping the Electrodes Moist
It is very important that the pH/ORP electrode is kept moist when not in use. This is
achieved by always fitting the storage cap, which incorporates a sponge that should be
soaked in a special storage solution.
The sponge within the storage cap should be moistened with a few drops of pH
Electrode Storage Solution each time it is removed and replaced.
If a pH/ORP
electrode is inadvertently allowed to dry out, it must be re-hydrated by soaking in storage
solution for at least one hour prior to use.