10.4. Calibration Error Messages
If the Aquameter
®
detects a problem with either the AP-5000 or the calibration solution
during the calibration procedure, an error will be indicated. The chart below shows the
possible errors and how to correct them.
Error
Message
Problem
Action
REPLACE DO
CAP
Full re-calibration required or Optical DO
Cap needs replacing
See
note below.
BATTERIES
TOO LOW
Battery Voltage is too low for reliable
calibration
Replace the batteries
NO PROBE
RESPONSE
The Probe is not responding
Check connections / cycle power
READINGS
UNSTABLE
Readings did not stabilise within the
expected period
Top up / replace the RapidCal
OUT OF CAL
RANGE
Readings are outside calibration limits (can
be caused by low level / incorrect calibration
solution). Or the Probe Sleeve is not fitted
Top up / check calibration
solution is correct type. Ensure
the Probe Sleeve is fitted
OUT OF TEMP
RANGE
Temperature is outside 5
C – 40
C limit
ISE differential calibration rules not met.
Warm / cool the RapidCal
CAL ZERO
FIRST
You are trying to calibrate an upper
calibration point on an electrode without
calibrating the zero point first.
Calibrate the zero point first,
then without switching the
Aquameter off, calibrate the
upper point.
If the 'REPLACE DO CAP' error occurs during Optical DO Zero calibration, this usually
indicates that the DO Cap needs replacing. Perform a full DO calibration first at DO Zero
then at 100% DO. If that does not cure the problem, replace the DO Cap ( see
in section 14)
.
If the corrective actions shown above for 'READINGS UNSTABLE' or 'OUT OF CAL
RANGE' errors do not work, thoroughly clean the Probe and try again. If the 'OUT OF CAL
RANGE' error persists, reset the calibration values to Factory Defaults then try again.
If the 'OUT OF CAL RANGE' error persists when calibrating EC, check you are using the
correct EC Calibration Standard and that the sleeve is fitted.
If the 'OUT OF CAL RANGE' error persists when calibrating pH, check you are using the
correct pH Calibration Standard for the calibration point selected.
If the 'OUT OF TEMP RANGE' error persists when carrying out a three point ISE
calibration, check your solution temperatures are within the specified limits with respect to
each other.
Remember:
The Probe sleeve and End Cap form an integral, working part of the
Probe’s optical and EC measurement system, and MUST be fitted during calibration
and measurement for correct operation.
If you try to calibrate the Probe without the
sleeve fitted, you may get an error message.