Working Standard
User’s Guide
WS-UGFW619-EN
www.appliedp.com
54
Always Show
..............
ON
/
OFF
. When enabled (
ON
), the Test results details screen will
automatically appear after end of meter error test (after registering
predefined number of impulses). It is the same like pressing
F3
key
after test done.
SNAP SWITCH REG. MODE
Each impulse
..............
ON
/
OFF
. When enabled (
ON
) the required number of pulses
(parameter
No. of impulses
in Test Settings screen 1/2) has to be
registered. When disabled (
OFF
) only the 0
st
and last pulse has to be
registered, thus n+1 pulses has to be counted. Pulses registration is
carried out by Snap Switch accessory or
F3
key in Meter Error Test
screen when registration mode is set to value
SNAP.S
.
ENERGY TESTS
..................... Settings of energy tests
Mode
........................... Mode of energy testing:
Time
or
Energy
Time
............................ when this time expires, the test procedure will terminate
automatically
Energy
......................... when this energy amount is measured, the test procedure will
terminate automatically
Energy unit
................. Definition of energy unit:
kWh
or
MWh
Energy type
................ Definition of energy type:
Active
,
Reactive
or
Apparent
AUX control
................
ON
/
OFF
. It is the AUX output controlled by energy test. When
enabled (
ON
) the LED is shining during the energy test and AUX
output is not controlled by energy output.
MAXIMUM DEMAND
.............. Settings of Maximum Demand Test
Period
.......................... duration (in minutes)
Subperiods
................. number of sub-periods if applicable (if not, use 0)
Window
....................... type of Maximum Demand Test time period (window):
Sliding
- continuous registering in last X minutes (defined in Period
parameter) … this means that new maximum demand value is
available every minute
Fixed
-
block registering from last full block with defined duration …
this means that new maximum demand value is available after X
minutes (defined in Period parameter)