47
2.
Calibration of the Ultrameter
III
gain Factor for User mode is
not available when the calibration linc has been established.
The other calibration functions (i.e., Temperature Compensation
%/C settings and TDS Ratio settings) are still intact. To perform
a calibration of the User mode as described in User Calibration
Conductivity/TDS, pg. 28, the User mode Linc should be canceled.
See above, “Canceling User mode calibration “
linc
””.
3.
Once a “
linc
” has been established for User mode, the “
linc
”
will apply to all measurement modes using User solution
selection (i.e., TDS/User, Cond/User or Res/User).
Xvi. bluDock™ wireless Data transFer instrUctiOns
nOte:
Bluetooth® is a registered trademark of Bluetooth SIg. The
bluDock Bluetooth module is a registered Bluetooth device.
Requires Myron L bluDock™ accessory package, Model # BLUDOCK.
Package includes Ultrameter
III
hardware modification that allows the
unit to communicate wirelessly with a personal computer configured for
wireless device communication. Package also includes U2CI software
application that will operate on Windows XP, Vista and 7*, and Macintosh
OSX** based computer systems and Bluetooth USB adapter (dongle)
for computers that do not have Bluetooth capability.
A. Software Installation
1.
Place Myron L Ultrameter
III
U2CI Installation CD v2.0.0 &
later into your computer or download U2CI application from the
Myron L website:
http://myronl.com/main/U2CI_Application_DL.htm
2.
Upon opening, select the folder for your operating system.
3.
Install U2CI application. See detailed installation instructions
on CD or Myron L website:
http://myronl.com/main/U2CI_Application_DL.htm
4.
Additional drivers may be required. See our website for the
latest information.
B.
Hardware Setup
For a computer without Bluetooth® capability:
If you don’t have the dongle that came with the BLUDOCK, one can be
ordered separately from the Myron L Company. Order Model # BDDO.
Plug in your dongle and install per manufacturer’s instructions.
Distributed by Applied Membranes, Inc. www.appliedmembranes.com
Summary of Contents for MYRON L COMPANY Ultrameter III 9PTK
Page 2: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 4: ...ii Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 73: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 75: ...Distributed by Applied Membranes Inc www appliedmembranes com ...