26
vii.
calibratiOn
A.
Calibration Intervals
generally, calibration is recommended about once per month with
Conductivity or TDS solutions. Calibration with pH solutions should
be checked twice a month. Calibration of ORP is not necessary (ref.
CALIBRATION INTERVALS TABLE
under Calibration Steps below).
B.
Rules for Calibration of the Ultrameter
III
1. Calibration Steps
The following table lists measurement functions and their corresponding
calibration requirements:
CALIBRATION INTERVALS TABLE:
Function
Kcl, nacl or 442
User
Cond
gain Only
Tempco, then gain
Res
Done in Conductivity
Done in Conductivity
TDS
gain Only
Tempco, Ratio, then gain
pH
7, acid and/or base
ORP
Zero set with pH 7
automatically
Alkalinity
gain Only
Hardness
gain Only
a. Starting Calibration
For Alkalinity and Hardness, calibration is begun by pressing
when the titration is complete and a value displays.
For Conductivity, TDS or pH, calibration is begun by pressing
while measuring Conductivity, TDS or pH. Measuring continues, but
the “
cal
” icon is on, indicating calibration is now changeable. The
reading is changed with the
and
keys to match the
known value.
The calibration for each of the 4 solution types may be performed in
either conductivity or TDS mode.
Once in CAL mode, the
key becomes an “ACCEPT” key for the
new calibration.
For pH, at each point, pressing
accepts the new calibration
value and steps you to the next adjustment.
Distributed by Applied Membranes, Inc. www.appliedmembranes.com
Summary of Contents for MYRON L COMPANY Ultrameter III 9PTK
Page 2: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 4: ...ii Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 73: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 75: ...Distributed by Applied Membranes Inc www appliedmembranes com ...