45
Follow the steps below to set either the KCl, NaCl or 442 calibration
factor to the User solution mode.
1.
Press measurement key desired to be “Linked”, i.e.,
,
or
.
2.
Place the Ultrameter
III
in User mode (ref. SOLUTION
SELECTION, pg. 22, for selecting the User mode).
3.
Press
arrow key until the menu “
linc
” appears (see
Figure 27).
4.
Press
key. The instrument will display “
sel
” and the
“
User
” Icon (see Figure 28).
Any additional display of KCl, NaCl or 442 icons indicates
a “Linc” between the User solution and the other solution
displayed.
5.
Press
or
keys to select a Standard Solution to be
linked to the User mode calibration constant (see Figure 29).
“
User
” is linked to “
Kcl
”.