Xviii. trOUbleshOOting chart
symptom
Possible cause
corrective action
No
display
, even though
measurement key pressed
Battery weak or not connected.
Check connections or replace battery. Ref. Battery Replacement, pg. 50.
Inaccurate
ph
readings
1. pH calibration needed. Ref. pH Cal.,
pg. 33.
2. Cross-contamination from residual pH
buffers or samples in sensor well.
3. Calibration with expired pH buffers.
1. Recalibrate instrument.
2. Thoroughly rinse sensor well.
3. Recalibrate using fresh buffers. Ref. pH Buffer Solutions, pg. 55.
No response to
ph
changes
Sensor bulb is cracked or an
electromechanical short caused by an
internal crack.
Replace pH/ORP sensor. Ref. Replacement pH/ORP Sensor, pg. 56.
Will not adjust down to
ph
7
pH/ORP sensor has lost KCl.
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 56).
ph
readings drift or respond slowly to
changes in buffers/samples or “
Fac
”
is displayed repeatedly
1. Temporary condition due to memory” of
solution in pH sensor well for long periods.
2. Bulb dirty or dried out.
3. Reference junction clogged or coated.
Clean and rejuvenate sensor (ref. Cleaning Sensors, pg. 50) and recalibrate. If no
improvement, replace pH/ORP sensor (ref. Replacement pH/ORP Sensor, pg. 56).
Unstable
conductivity/tDs/
resistivity
readings
1. Dirty electrodes.
2. Actual resistance is changing due to
atmospheric contamination.
1. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
2. Minimize test sample exposure to air by taking a flowing sample.
Ref. Measuring Resistivity, pg. 14.
Unable to calibrate
conductivity/
tDs
Film or deposits on electrodes
.
Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
resistivity
readings much lower than
expected
1. Contamination from previous sample or
from pH sensor well.
2. Carbon dioxide in test sample.
1. Rinse cell cup more thoroughly before measurement. Ensure pH cap is snugly in
place.
2. See Measuring Resistivity, pg. 14.
alkalinity
or
hardness
titration with
calibration
solution value not as
stated on bottle
1. Method error.
2. Dirty Cell Cup.
3. Dirty pipette.
4. Faulty pipette.
5. Contaminated calibration solution.
1. Repeat titration.
2. Clean cell cup and electrodes. Ref. Cleaning Sensors, pg. 50.
3. Clean pipette per manufacturer’s instructions.
4. Use a new pipette.
5. Use fresh calibration solution. Be sure to use a new tip for each type of solution
measured.
Low Reading
Slow or no response to
OrP
changes
ORP platinum electrode is dirty.
Clean ORP electrode with ORP conditioner. Ref. Cleaning Sensors, 2. pH/ORP
Sensor, pg. 50.
Distributed by Applied Membranes, Inc. www.appliedmembranes.com
Summary of Contents for MYRON L COMPANY Ultrameter III 9PTK
Page 2: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 4: ...ii Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 73: ...Distributed by Applied Membranes Inc www appliedmembranes com ...
Page 75: ...Distributed by Applied Membranes Inc www appliedmembranes com ...