b.
In the Run screen, click Start Run.
IMPORTANT!
Do not attempt to open the access door during the run. The
door is locked while the instrument is in operation.
Note:
Before starting the calibration, the instrument may pause (up to
10 minutes) to allow the heated cover to reach the required temperature.
8.
When the run is complete and the QuantStudio
™
Real-Time PCR Software
displays the Analysis screen, confirm the status of the calibration. The analysis
status indicates the success of the calibration, where passed indicates that the run
produced viable calibration data, and failed indicates that the run did not
produce data or that the data it collected is unusable.
Analysis status
Action
Passed
Enter comments in the Comments field, click Next.
Failed
Troubleshoot the failed calibration (see “Troubleshooting“ on
9.
Select the QC tab to review the quality check summary. The QC status indicates
the quality of the calibration data, where passed indicates that all wells produced
data that passed the quality check, and failed indicates that one or more wells
produced unusable data.
Note:
If the calibration failed, the QC tab can provide information regarding the
cause of the failure.
10.
Unload the array card:
a.
After the calibration, tap on the touchscreen to eject the array card.
Note:
If the instrument does not eject the array card, remove it as explained
in “Troubleshooting“ on page 72.
b.
Remove the array card from the instrument tray and discard it.
WARNING!
PHYSICAL INJURY HAZARD. During instrument
operation, the array card temperature can reach 100°C. Allow the
consumable to cool to room temperature before removing.
c.
Tap on the instrument touchscreen to close the instrument tray.
11.
In the Normalization Calibration screen, click Finish to complete the calibration,
then click Yes when prompted to save the results.
12.
Repeat the calibration procedure to perform the remaining normalization
calibration.
13.
(Optional) Click Print Report in the upper right corner of the screen to print a
summary of the calibration results for your records.
Chapter 2
Calibrate the instrument
Normalization calibration (TaqMan
®
Array Card only)
2
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
61