b.
In the Instrument Console, select your instrument from the list of
instruments on the network, then click Add to My Instruments.
Note:
You must add a system to your list before you can manage it.
c.
After the system is added to your list, select it, then click Manage
Instrument.
d.
In the Instrument Manager, click Maintenance, then click Calibration
Reminders.
2.
In the Calibration Reminders Setting table, configure the notification settings for
the calibrations in interest. For each calibration that you want to monitor:
a.
In the Expiry Interval column, enter the number of days that elapse before
the type of calibration expires on the system.
b.
In the Send a Reminder column, select the check box to configure the
QuantStudio
™
Real-Time PCR Software to email a reminder to perform the
calibration.
c.
In the Reminder Interval column, enter the number of days that elapse
before the software emails recipients a reminder to perform the calibration.
3.
In the Enter e-mail addresses for notifications field, enter the email address(es)
that you want to receive email notifications. Separate multiple email addresses
with commas (
,
).
4.
Click Apply to change the notification settings.
Appendix C
Networking
Monitor the QuantStudio
™
6 or 7 Flex Real-Time PCR System
C
122
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide