4.1.13. Ensure that the BEFORE, DURING, and AFTER gauges on Separator vessel gauges
and verify Separator pressure on the home screen reads default zero.
4.1.14. Remove the flexible metal line and connection pipe from the top of the separators.
Use two wrenches to prevent the NPT fittings from loosening in the separator cap.
Figure 25. Illustration of using two wrenches to remove flexible metal lines
4.1.15. Remove the yellow wire connected to the Separator #1 thermocouple.
4.1.16. Use the 5/8-in ratchet wrench to remove the high pressure sanitary clamps from the
top of separator #1.
4.1.17. Remove the cap from the top of separator.
4.1.18. Collect any available oil from the separator cap.
4.1.19. Use alcohol to clean the caps and orifice tube.
4.1.19.1. Separator 1 cap must be cleaned every run.
4.1.20. Use the supplied round squeegee or hard plastic scrapper to push any residual oil
from the sides of the separator #1 down to the bottom and in the collector cup.
BEFORE
AFTER
DURING