Updated 6/16/2016
Copyright Apeks LLC, 2014
34
Recommended Operating Parameters
4.1.74 Recommended default parameters are intended as a starting point and it is up to the
operator to determine best results for their operation. Please note that ambient
temperature effects parameters and the system will take an hour to equalize out.
4.1.75 PDC 3 Diaphragm Compressor
Subcritical:
Extractor Pressure: 1,200 psi
Extractor Temperature: 75F
Orifice Size: 22
Separator Pressure: 350-380psi
Separator Temperatures: +/- 5 degrees around 23F
Extraction Time: 3-4 hours per pound
Supercritical:
Extractor Pressure: 1,800 psi
Extractor Temperature: 100F
Orifice Size: 18
Separator Pressure: 350-380psi
Separator Temperatures: +/- 5 degrees around 64F
Extraction Time: 2-3 hours per pound
4.1.76 PDC 4 Diaphragm Compressor
Subcritical:
Extractor Pressure: 1,200 psi
Extractor Temperature: 75F
Orifice Size: 37
Separator Pressure: 350-380psi
Separator Temperatures: +/- 5 degrees around 21F
Extraction Time: 2-3 hours per pound
Supercritical:
Extractor Pressure: 1,800 psi
Extractor Temperature: 100F
Orifice Size: 29
Separator Pressure: 350-380psi
Separator Temperatures: +/- 5 degrees around 69F
Extraction Time: 1-2 hours per pound
Summary of Contents for 2000 PSI Series
Page 49: ...Updated 6 16 2016 Copyright Apeks LLC 2014 49 ...
Page 50: ...Updated 6 16 2016 Copyright Apeks LLC 2014 50 Appendix C CO2 Phase Diagram ...
Page 51: ...Updated 6 16 2016 Copyright Apeks LLC 2014 51 ...
Page 52: ...Updated 6 16 2016 Copyright Apeks LLC 2014 52 Appendix D Training Process ...