light turns on.
Press
【
ENTER
】
〖
01800
〗
Based on the first load, it will display the
minimum weight required for the second
load. From our example, it may display
1800.
Press
【
TARE
】
and
【
ZERO
】
or
【
DATE
】
〖
03000
〗
The LOAD2 shall be 3000 (from our
example). Press
【
ZERO
】
to adjust
digit’s value and press
【
TARE
】
to move
to next digit. Change the value here into
Input 3000 and wait until STB light turns
on.
Press
【
ENTER
】
〖
LoAd3
〗
Hanging the third load.
Press
【
ENTER
】
〖
03600
〗
Mention your Load3 must exceed 2600kg
Press
【
TARE
】
and
【
ZERO
】
or
【
DATE
】
〖
05000
〗
From our example LOAD3 shall be
5000kg. Change the value here into 5000
and wait until STB light turns on
Press
【
ENTER
】
〖
------ -
〗
〖
End
〗
〖
5000
〗
Press
【
MENU
】
or
【
MODE
】
〖
ESC
〗
Press Menu or Mode multiple times until
it displays Esc then press enter to exit
calibration.
First load: LOAD1 must be larger than 20% of its FS capacity and
LOAD1<LOAD2<LOAD3. Between LOAD1 and LOAD2 and LOAD2 and LOAD3,
the interval must be greater than 20% of its FS capacity.
8.
Display Illustration
DISPLAY
ILLUSTRATION
DESCRIPTION
〖
noSIg
〗
Without wireless signal
Distance too far.
〖
SEtUP
〗
Enter into parameter setting
〖
UAdJ
〗
Enter voltage calibration
〖
LoAd
〗
Calibration point
〖
¯ ¯ ¯ ¯ ¯
〗
Exceed high limit
Tare weight cannot exceed full
capacity
〖﹍﹍
〗
Exceed low limit
Tare weight cannot be negative
〖
-----
〗
Waiting stable
〖
Err10
〗
Weight less than Min. Capacity
Cannot accumulate the value
〖
Err11
〗
Accumulated times overflow
Cannot accumulate after 30
20
|
ANYLOAD OCSD Wireless Dynamometer Operations Manual (V1904)