OPERATION
DISPLAY
DETAILS
Press
【
MENU
】
or
【
MODE
】
(insert to PC)
〖
LOCAL
〗
Press
【
ZERO
】
or
【
DATE
】
〖
CLInt
〗
Press
【
ENTER
】
〖
u 1.03
〗
Displays the Current indicator Software
version
〖
88888
〗
〖
Ert
〗
〖
CH=E3
〗
Displays the current indicator channel E3
〖
U=X.XX
〗
Displays the current indicator voltage
which is X.XX
〖
SETdC
〗
Press
【
MENU
】
or
【
MODE
】
〖
SEtrF
〗
Dynamometer’s RF parameter setting
Press
【
MENU
】
or
【
MODE
】
〖
SEt3P
〗
Enter into a 3 point calibration
Press
【
ENTER
】
〖
CAL 3
〗
In these instructions, we will set 5000kg
scale as an example. For the 5000kg
scale, the recommended calibration
weights are LOAD1: 1500kg, LOAD2:
3000kg, LOAD3: 5000kg
Press
【
ENTER
】
〖
CALSP
〗
Zero point calibration. Make sure the
stable light is on before running this
procedure
Press
【
ENTER
】
〖
LoAd1
〗
At this stage you will load the LOAD1
which is 1500kg (from our example). First
load must exceed 20% of its capacity.
Press
【
ENTER
】
〖
05000
〗
It will display FS capacity you set in
parameter settings.
Press
【
TARE
】
and
【
ZERO
】
or
【
DATE
】
〖
01000
〗
Change the value into 1500.
Press
【
ZERO
】
to adjust digit’s value
and press
【
TARE
】
to move to next digit.
Wait until STB light turns on.
Press
【
ENTER
】
〖
LoAd2
〗
Hanging the second load LOAD2 which is
3000kg (from our example). Second load
must be at least 20% of its capacity
greater than the first load. Wait until STB
19
|
ANYLOAD OCSD Wireless Dynamometer Operations Manual (V1904)