indicator will alarm when load is
smaller than the value.
Press
【
ENTER
】
to
confirm
〖
02000
〗
The screen should display 02000 if
parameter is set to 2000.
Press
【
ZERO
】
or
【
DATE
】
and
【
TARE
】
to
change value
〖
01000
〗
Set “1 HI” or “1
LO” Value. e.g.
1000kg
Press
【
ENTER
】
〖
SP2
〗
Set point 2
Press
【
ENTER
】
to
confirm
〖
2 OFF
〗
The screen will display 2 OFF, if this
parameter was set. It will display the
set value.
Press
【
ZERO
】
or
【
DATE
】
to
choose
〖
2 HI
〗
or
〖
2 LO
〗
There are 3 choices: 2 OFF means you
will not set set point; 2 HI means the
indicator will alarm when load exceeds
the value you set; 2 LO means the
indicator will alarm when load is
smaller than the value
Press
【
ENTER
】
to
confirm
〖
01200
〗
The screen should display 01200 if
parameter is set to 1200.
Press
【
ZERO
】
or
【
DATE
】
and
【
TARE
】
〖
03000
〗
Set “2 LO” or “2 HI” Value.
e.g.3000kg, input 3000
Press
【
ENTER
】
〖
g=
〗
Acceleration due to gravity
Press
【
ENTER
】
,
【
ZERO
】
or
【
DATE
】
and
【
TARE
】
to
change value
〖
9.7930
〗
The screen will display the default or
the calibrated (e.g 9.7930). User can
change it according to local area gravity
value. The value range is 9.783 - 9.832.
Press
【
MENU
】
or
【
MODE
】
key
〖
End
〗
〖
0
〗
Confirm above parameter setting. Exit
parameter setting program.