Below is the recommended division and decimal position of the scale according
to its full scale capacity:
CAPACITY/DIVISION FS
(
CAP
)
Id (DIVISION)
Pt
(
Decimal
position
)
1000kg/0.5kg
10
05
1
2000kg/1kg
02
01
0
3000kg/1kg
03
01
0
5000kg/2kg
05
02
0
10000kg/5kg
10
05
0
15000kg/5kg
15
05
0
20000kg/10kg
20
10
0
30000kg/10kg
30
10
0
50000kg/20kg
50
20
0
7.3 Calibration
User must set all parameters before doing calibration.
7.3.1 One Point Calibration
OPERATION
DISPLAY
DETAILS
Turn on
dynamometer and
indicator
〖
0
〗
In these instructions, we assume we have
5000kg dynamometer. Make sure it is
displaying
〖
0
〗
before starting the
calibration. At this stage you can
hang tare weight ( if it is required)
then reboot the dynamometer to get
zero reading.
Press
【
MENU
】
or
【
MODE
】
8 times
〖
CLIbr
〗
Access calibration menu
Press
【
ENTER
】
key
〖
CAL 1
〗
When
〖
CAL 1
〗
is displaying press
enter to start the one point
calibration
Press
【
ENTER
】
key
〖
CALSP
〗
In this stage we have to run the zero
calibration. Make sure the scale has
17
|
ANYLOAD OCSD Wireless Dynamometer Operations Manual (V1904)