calibration, set Cd=00 to achieve high
accuracy. After calibration, set Cd back
to Cd=11 again. Generally choose
Cd=12.
Press
【
ENTER
】
to confirm
〖
LL=2
〗
If Dynamometer was already calibrated,
it will display the calibrated value for
filter parameter (e.g LL=2).
Press
【
ZERO
】
or
【
DATE
】
to choose
〖
LL=1
〗
LL: filter parameter has 6 options from
smallest to biggest. Before calibration,
set LL=0, after calibration, set LL=1,
generally choose LL=1.
Press
【
ENTER
】
to confirm
〖
Un=0
〗
If Dynamometer was already calibrated,
it will display the calibrated value for
the measurement unit (e.g Un=0).
Press
【
ZERO
】
or
【
DATE
】
to choose
〖
Un=0
〗
Set Un=0 for kg, Un=1 for lb, Un=2 for N,
Un=3 for KN, Un=4 for t, Generally,
choose Un=0.
Press
【
ENTER
】
to confirm
〖
OFF 1
〗
If Dynamometer was already calibrated,
it will display the calibrated value for
auto off function (e.g OFF 1).
Press
【
ZERO
】
or
【
DATE
】
to choose
〖
OFF 2
〗
Auto Off has 3 options: 0: turn off by
hand; 1: if no operation, the screen
will display ----- after 15minutes to
save power; 2: if no operation, the
screen will display ----- after 15minutes,
and turn off automatically after
60minutes. Generally, choose OFF=2.
Press
【
ENTER
】
〖
SP1
〗
Set point 1
Press
【
ENTER
】
to
confirm
〖
1 OFF
〗
The screen will display 1 OFF, if this
parameter was set. It will display the
set value.
Press
【
ZERO
】
or
【
DATE
】
to
choose
〖
1 HI
〗
or
〖
1 LO
〗
There are 3 choices: 1 OFF means you
will not set set point; 1 HI means the
indicator will alarm when load exceeds
the value you set; 1 LO means the
15
|
ANYLOAD OCSD Wireless Dynamometer Operations Manual (V1904)