scale will be set to 6000kg capacity
scale. If set to FS=10 then the capacity
will be 10000kg.
Press
【
ZERO
】
or
【
DATE
】
〖
FS=05
〗
Assuming we are setting a 5-ton scale,
set FS=05. Here are the available
capacities that can be set:02 / 03 / 05 /
06 / 10 / 15 / 20/30/50/60/75/80.
Press
【
ENTER
】
to
confirm
〖
Id=02
〗
The screen may display Id=02. This
maybe the division set during previous
calibration.
Press
【
ZERO
】
or
【
DATE
】
to
choose
〖
Id=02
〗
Assuming we will set the 5-ton scale
into 2kg division, so, we will set ld=02.
Here are the available options: 01 / 02 /
05/10/20 circularly.
Press
【
ENTER
】
〖
Pt=0
〗
The screen may display Pt=0. This
indicates the decimal point position. We
will set it to Pt=0 for the 2kg division.
Press
【
ZERO
】
or
【
DATE
】
to
choose
〖
Pt=0
〗
Choose decimal position. The options
are 0/1/2/3 . 0=xxxxx, 1=xxxx.x,
2=xxx.xx, 3=xx.xxx. For Cap.5000kg ,we
will set Pt=0
Press
【
ENTER
】
to confirm
〖
Ab=24
〗
Zero range settings. A: zero range by
hand (manual); B: automatic zero range.
There are 6 options: 0=0%F. S;
1=2%F.S; 2=4%F.S; 3=10%F.S; 4=20%F.S;
5=50%F.S. Generally do not change this
parameter.
Press
【
ENTER
】
to confirm
〖
Cd=11
〗
If Dynamometer was already calibrated,
it will display the calibrated value for
zero tracking range (e.g Cd=11).
Press
【
ZERO
】
or
【
DATE
】
to choose
〖
Cd=12
〗
C: zero-tracking range. There are 6
options; 0=0d; 1=0.5d; 2=1d; 3=1.5d;
4=2d; 5=2.5d
d: display speed. There are 3 options:
0=slow; 1=average; 2=fast. Before
14
|
ANYLOAD OCSD Wireless Dynamometer Operations Manual (V1904)