Data too big after changing
units
Check decimal setting in
primary and secondary
units. Refer Section 9.2.2
regarding F2 Menu
– – E 1 – –
Incorrect operating
parameters
Check parameters
according to Section 9
– – E 2 – –
A/D exchange error S
Check hardware by
qualified personnel
– – E 3 – –
Data reading error
Check hardware by
qualified personnel
– – E 4 – –
A/D transfer start up error
Check hardware by
qualified personnel
– – E 5 – –
Count item average weight
too small
Check scale range or
increase sampling weight
according to Section 11
– – E 6 – –
No average weight set for
count items
Refer to Section 11 to set
average weight
– – E 7– –
Load cell input signal>20mV
Check load cell and
connecting cables
– – EL – –
Load cell input signal<-4mV
Check load cell and
connecting cables
– – OF – –
Load value>F1.5 Set value
Reduce load on scale
15.2
Software Version and Default Configuration Parameters
Fig. 15.2.1 F7(Ver) Menu
52
|
ANYLOAD 805TS & 805BS Series Weighing Indicator Operations Manual (v1703)
15.
Appendices
15.1 Error Messages
Error Message
Description
Solution
– – E 0 – –