3 (NONE)
NONE is selected, Tare removal
can be done at any weighing
mode.
For NTEP and OIML, a new Tare
can be acquired even when
there is a stored Tare.
For CANADA, a new Tare can be
acquired after the stored Tare is
removed.
For NONE, NTEP and CANADA,
when the current weight is
within the specified zero range,
zero scale can be performed
irrespective it is in Gross or Net
mode.
For OIML, zero scale can only be
performed when it is in Gross
mode and Tare can be removed
when it is in Net mode.
F6.2
(
P1
)
This function is not
available in 805BS-TN-B
Lite Version model
Number
Specifies fixed value 1. Refer to
Section 12 for Relay Output
setup.
F6.3
(
P2
)
This function is not
available in 805BS-TN-B
Lite Version model
Number
Specifies fixed value 2. Refer to
Section 12 for Relay Output
setup
F6.4
(Password)
For 805TS-B-17, 805BS-
B-17 & 805BS-TN-B
Number
The password is required when
entering or exiting configuration
mode. (refer to Section 15.2.3)
F6.5
(User Overload)
For 805TS-B-17, 805BS-
B-17 & 805BS-TN-B
Number
When net weight exceeds user
overload value, al
arm will start
beeping.
Note: If F6.4 = 0 & F6.1 = 3 (none), SW2 switch is still used to enter and exit
calibration
menu. If F6.4 ˃ 0 & F6.1 = 3 (none) and SW2 switch is ON
, password
is required for entering and exiting calibration menu.
31
|
ANYLOAD 805TS & 805BS Series Weighing Indicator Operations Manual (v1703)