![Anyload 805BS-B-17 Operation Manual Download Page 47](http://html1.mh-extra.com/html/anyload/805bs-b-17/805bs-b-17_operation-manual_2948860047.webp)
(
1
)
Turn on the SW2 switch to set indicator to the Configuration mode.
Display shows “F1”.
(
2
)
Press until display shows “F7” (see Fig. 13.2.1). Press
to go to
F7.1 Menu. Display shows “F7.1”
(
3
)
When display shows “F 7.1”, press
. Display shows the current
consecutive number “nnnnnn”. Edit numbers according to Fig. 4-2.
(
4
)
After editing, press
to return to F7.2 Menu. Display shows “F7.2”
(
5
)
When display shows “F7.2”, press
. Display shows “nnnnnn”. Edit
numbers according to Fig. 4-2.
(
6
)
After editing, press
to return to F7.3 Menu. Display shows “F7.3”.
(
7
)
Turn off the SW2 switch to exit from Calibration mode and enter into
Weighing Mode.
13.5 Date and Time Setting
(
1
)
Follow points (1) and (2) to enter into F7 Menu.
(
2
)
When display shows “F7.1”, press
until display shows “F7.4” (see
Fig. 13.4). Press
to enter into Date setup mode and display shows
“DD.MM.YY”.
date.
(
3
)
When date setup is complete, press
to return to F7.5 Menu.
Display shows “F7.5”.
(
4
)
When display shows “F7.5”, press
and display shows time
“HH.MM.SS”. Follow Fig. 4
-2 to set the current time.
(
5
)
When complete, press
to return to F7.6 Menu. Display shows
“F7.6”.
(
6
)
Turn off the SW2 switch to exit from Calibration mode and enter into
Weighing Mode.
14.
Serial Communication
Indicator has the following two serial communication modes:
o
Continuous transmission
o
Transmission upon request (from an external PC)
46
|