2)
Press until the display shows “F3” (see Fig. 10.2). Press
to go to
zero calibration mode. Display shows “F3.1”.
3)
When display shows “F3.1”
, p
ress
to zero calibrate the indicator.
Display will show “CAL” while calibration is in progress. When completed
,
the A/D count for the zero calibration is shown. You may record this A/D
count. This A/D count ( usually a 6-digit count ) can be used if
recalibrating the indicator using the Digit Calibration Method. Do not
adjust this value. Then press
again to save the zero calibration value
and go to the next menu (display shows “F3.2”).
4)
When display shows “F3.2”, place test weight on the scale and press
to show the test weight value. Follow Fig. 10.2 to input the test weight
value, then press
to save the value and go to the next menu. Display
shows “F3.3”.
5)
When display shows “F3.3”
, p
ress
to span calibrate. Display shows
“CAL” while calibration is in progress. When completed
, the A/D count
for the span calibration is shown. You may record this A/D count. This
A/D count ( usually a 6-digit count ) can be used if recalibrating the
indicator using the Digit Calibration Method. Do not adjust this value.
Press
again to save the span calibration value and go to the next
menu. Display shows “F 3.4”.
6)
F3.4 Menu is used to remove a calibration offset when hooks or chains
are used to hang the test weights.
When display shows “F3.4”, there are 2 options:
•
If no other apparatus are used to hang the test weights during
calibration, remove the test weight and press
to return to F4
Menu. Display shows “F4”
•
If hooks or chains are used during calibration, remove these and
the test weights from the scale. With all weight removed, press
to re-zero the scale. This function adjusts the zero and span
calibration values. Display shows “CAL” while zero and span
calibrations are adjusted. When completed
, the adjusted A/D
count for the zero calibration is shown. You may record this A/D
count. This A/D count ( usually a 6-digit count ) can be used if
recalibrating the indicator using the Digit Calibration Method. Do
not adjust this value. Press
to save the value and to return to
35
|
ANYLOAD 805TS & 805BS Series Weighing Indicator Operations Manual (v1703)