6.3. SUCTION TUBES
Suction starts when lifting the suction tube holder off its support.
To adjust suction power, use the lever (
a
) located on the suction tube
holder handle.
The dental unit is equipped with the V.D.S. system that allows the drying
of the suction line using an automatic stop delay (about 2 seconds).
For flushing procedure, see paragraph 7.6.
Removing the suction tubes.
Always wear goggles and gloves to prevent contact with
infected material when removing the suction tubes.
Remove the suction tubes from the conveyor fittings by turning and pulling
the tube fitting.
Detach the suction tubes from the holders by turning and pulling the tube
fitting.
Never directly grasp the suction tube.
Sterilization.
• Suction tube holder terminals: steam autoclave up to 135° C following the instructions for the use of the device.
• Suction tubes: soak to cold-sterilize.
With the tubes, never use procedures where temperature exceeds 55° C.
Flushing the suction tubes
As the dental units may be equipped with different suction systems (liquid or wet ring, air) carefully follow the instructions provided by the suction system
manufacturer when disinfecting the system regarding the product to be used, times and directions for use.
For cleaning of the suction system, it is recommended to use STER 3 PLUS (CEFLA s.c.) diluted in a 6% solution (equivalent to 60 ml
of product in 1 litre of water).
Maintenance.
Periodically lubricate the O-rings of the suction tube holder terminals (see Paragraph 9.4.) using
S1-Protective Lubricant for O-Rings
(CEFLA s.c.).
Note about biocompatibility.
Only use suction tubes supplied with the dental unit and original spare tubes.
The suction tubes must comply with Biocompatibility Standard ISO 10993.
ISOLITE suction tube.
For ISOLITE suction tube operation, please refer to the specific use instructions given by the manufacturer.
Summary of Contents for A5 CONTINENTAL
Page 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Page 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Page 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Page 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Page 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Page 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Page 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Page 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Page 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Page 99: ......
Page 100: ...www cefla com...