A5 - A6 PLUS - L6
10
EN
1.6. STERILISATION
Every instrument supplied is NOT STERILE and must be sterilised in steam autoclave (max. 135°C) before use, without performing any type of chemical
sterilization.
Sterilization must be performed using appropriate packaging materials, verified during the sterilization process validation.
It is recommended to sterilize in steam autoclave (moist heat) through a cycle with pre-vacuum (forced air removal).
Autoclaves must comply with and be approved and maintained in accordance with the requirements set forth by the Standards EN 13060 (or ANSI/AAMI
ST55), EN ISO 17665-1 and ANSI/AAMI ST79.
Below are the minimum recommended parameters for the sterilization of reusable medical devices, that have been approved in order to ensure a sterility
assurance level (SAL) equal to 10^-6:
•
Cycle type:
with pre-vacuum (Pre-vac).
•
Method
: "overkill" moist heat sterilization according to Standard ISO 17665-1.
•
Minimum temperature:
134° C (273° F) for heat-resistant materials (instruments and metal handpieces, etc.);121°C (250°F) for thermolabile materials
(rubber items, etc.).
•
Minimum exposure time (1)
:
4 minutes (at 134°C), 20 minutes (at 120°C).
•
Minimum drying time (2)
: defined to ensure compliance with the requirements of Standard EN 13060 (or ANSI/AAMI ST55).
(1)
Exposure time: time during which the load and the entire chamber are kept at a temperature higher than the sterilization one.
(2)
Drying time: time during which steam is removed from the chamber and the chamber pressure decreases to allow the condensate to evaporate from
the load through a prolonged emptying or thanks to the supply of hot air or other gases then removed.
The drying time varies according to the load configuration and the packaging type and material.
2. DESCRIPTION OF THE EQUIPMENT
2.1. IDENTIFICATION PLATES
Data given on plate:
• Name of manufacturer.
• Name of equipment.
• Rated voltage.
• Type of current.
• Rated frequency.
• Maximum current or power absorbed.
• Serial number.
• Date of manufacture.
Nameplate position.
Dental units model:
A5 CONTINENTAL
A5 INTERNATIONAL
A5 SINGLE WATER UNIT
A6 PLUS CONTINENTAL
A6 PLUS INTERNATIONAL
A5 CART
Dental units model:
A5 SINGLE CART
Summary of Contents for A5 CONTINENTAL
Page 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Page 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Page 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Page 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Page 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Page 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Page 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Page 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Page 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Page 99: ......
Page 100: ...www cefla com...