A5 - A6 PLUS - L6
EN
65
• Fit the tank back in place by turning it counter-clockwise.
Make sure that tank is duly fastened.
• Touch the button (
T
) to re-enable the S.H.S. system.
Check that icon (
A
) on DISPLAY turns on.
If you are going to be absent from the surgery for long periods of time (holidays), completely empty the tank (a) before leaving. It is
advisable to empty it by calling water to the cup (use the cup supplied), until air comes out of the nozzle.
Cleaning and disinfection of the tank.
It is recommended to periodically disinfect (at least once a month) the tank only, using as disinfectant Peroxy
Ag+ Cefla or 3% Hydrogen Peroxide
(Oxygenated Water, 10 volumes), acting as follows:
• Remove tank from dental unit and drain it completely.
• Fill tank to the brim with the disinfectant.
• Allow the disinfectant to remain in the tank for at least 10 minutes.
• Drain disinfectant from the tank.
• Rinse tank with distilled water.
• Fill tank with distilled water, if necessary charging it with the substances indicated above.
• Fit tank back into its seat inside dental unit.
7.2.1.
MANUAL S.H.S. SYSTEM (applied on A5 SINGLE CART model only)
System features a tank (
a
) for distilled water.
The tank has a total capacity of 1.8 litres.
Tank supplies:
• The sprays of all the instruments present on dentist's and assistant's
boards.
• The cup filling system.
• The water quick coupling (if available).
A by-pass toggle (
b
) allows disabling the system in case you want to
supply instruments with mains water.
This system allows performing a disinfection cycle of instrument spray
ducts using oxygenated water (see paragraph 7.2.2.).
Carry out a disinfection cycle at the end of every working
day.
Filling the tank.
Once tank min. level is reached (approx. 500 cc), fill it as follows:
• Switch the toggle (
c
) to "CLOSE AIR PRESSURE” position.
• Remove tank (
a
) by turning it clockwise.
During this operation the pressurised air present inside the tank will be automatically bled to the outside.
• Pour distilled water inside the tank until reaching the max. level.
Use distilled water only. For enhanced hygiene, it can be added with 600 parts per million (ppm) of hydrogen peroxide using 20 ml
of Peroxy Ag+ per litre of distilled water, or oxygenated water (20 ml of 3% oxygenated water per litre of distilled water).
• Fit the tank back in place by turning it counter-clockwise.
• Switch the toggle (
c
) back to "OPEN AIR PRESSURE" position.
•
Make sure that the tank is correctly secured in place before switching toggle (c) to "OPEN AIR PRESSURE" position.
•
If you are going to be absent from the surgery for long periods of time (holidays), completely empty the tank (a) before leaving.
Summary of Contents for A5 CONTINENTAL
Page 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Page 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Page 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Page 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Page 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Page 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Page 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Page 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Page 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Page 99: ......
Page 100: ...www cefla com...