A5 - A6 PLUS - L6
48
EN
Operation.
.
•
The instrument is supplied non-sterile.
•
Before use, disinfect the lamp handle. The fiber optic and
the eye protection can be sterilized in a steam autoclave
up to 135°C.
• Put the fiber optic (
c
) in its housing until it clicks.
• Attach the curing light handpiece to the end of its power hose and tighten
the ring nut (
e
).
• Take the lamp out of its housing on the assistant’s board or dentist's
board.
Instrument activation is highlighted by the relevant managing
screen appearing on the DISPLAY.
• Turn the front of the lamp and/or fiber optic to the position most suitable
for curing (wand, gun or intermediate positions).
• Use the MODE button to select the desired cycle as previously directed
(the selected cycle is always indicated by the illuminated LED).
The curing light has a permanent memory therefore the last
cycle used will always be present the next time it is used.
• Place the fiber optic in the position required for curing.
The fiber optic should be placed as close as possible to the
material to be cured without touching it.
• Press button START to start the cycle.
Operating mode: work 2 consecutive cycles, rest 60
seconds.
When a programmed cycle is activated, the LEDs (1, 2, 3, B,
R, L) indicate the time that elapses (in multiples of 5 seconds)
and turn off every 5 seconds of operation.
The curing light comes with a beep that BEEPS when the cycle
starts, BEEPS every 5 seconds of operation and lastly BEEPS
twice at the end of the work cycle.
• Allow light emission to stop by itself. However, it can be stopped at any
time by simply pressing the START button again.
•
The curing light is equipped with a system that signals
malfunctions by illuminating the LEDs in different
combinations (see next paragraph).
•
The curing light is equipped with a thermal cut-out.
Summary of Contents for A5 CONTINENTAL
Page 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Page 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Page 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Page 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Page 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Page 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Page 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Page 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Page 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Page 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Page 99: ......
Page 100: ...www cefla com...