6-8 Troubleshooting Test – 3739A Power Supply DC Check
Troubleshooting
6-8
PN: 10410-00306 Rev. F
VectorStar ME7838 Series MM
6-8
Troubleshooting Test – 3739A Power Supply DC Check
This procedure verifies that the expected DC voltages from the power supply is present in the 3739A Test Set.
Equipment Required
•
Digital Multimeter
Procedure
1.
Turn off the 3739A Test Set and unplug the AC power cord.
2.
Remove the top cover of the 3739A Test Set.
3.
Connect the Test Set to AC power and turn the unit back on.
4.
Use a digital multimeter to measure the DC voltages at the test points stated in
on the Test Set
Bias Control PCB Assembly and verify if they are at the expected level. Refer to
for locations of test points.
5.
If any of the voltages is not present, replace the power supply.
Table 6-2.
Power Supply Expected DC Voltages
Test Point
Common
Expected Voltage
E17
TP1
5 V
E18
TP1
12 V
E19
TP1
12 V
E20
TP1
12 V
Summary of Contents for VectorStar ME7838 Series