Adjustment
5-3 ALC Level Calibration
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
5-17
24.
After calibration is complete, disconnect the Anritsu SC7770 power sensor from the Millimeter-Wave
Module at Port 2.
25.
Connect the Keysight W8486A power sensor to the Test Port of the Millimeter-Wave Module at Port 2.
26.
Change the
Freq Range
to
Bnd2:70G-125GHz.
27.
Select
Port 2 ALC Cal
.
28.
Click the
Start Cal
button to start the calibration. Wait until the calibration is complete.
Port 2 IF Leveling Calibration
29.
Select
Back
and then
IF Leveling.
30.
Select
Port 2 ALC Cal
.
31.
Click the
Start Cal
button to start the calibration.
32.
After calibration is complete, disconnect the Keysight W8486A power sensor from the Millimeter-Wave
Module at Port 2.
33.
Change the Freq Range to
Bnd1:54G-70GHz
.
34.
Connect the Anritsu SC7770 power sensor to the Test Port of the Millimeter-Wave Module at Port 2.
35.
Select
Port 2 ALC Cal
.
36.
Click the
Start Cal
button to start the calibration.
37.
The ALC Calibration is now complete.
38.
Install the waveguide adapters back to the Millimeter-Wave modules.