Adjustment
5-2 LO Level Calibration
VectorStar ME7838 Series MM
PN: 10410-00306 Rev. F
5-3
7.
Disconnect the LO cable from the
Port 1 LO
connector of the 3739A, 3739B or 3739C Test Set.
8.
Connect the Coaxial Directional Coupler with the Adapter to the
Port 1 LO
connector of the Test Set.
9.
Connect the LO cable to the RF Out port of the Coaxial Directional Coupler.
10.
Connect the power sensor to the Coupling Port of the Directional Coupler.
11.
On the MS464xx VNA, select
System
and then
Diagnostics
.
12.
The
Diagnostics Access
dialog box appears providing an entry field to enter the diagnostics access
password as shown below in
13.
Enter the password
CajaNueva
in the
Password
field and click
OK
.
14.
The
Diagnostics Menu
appears as shown in
Figure 5-2.
DIAGNOSTICS ACCESS Dialog Box
Figure 5-3.
DIAGNOSTICS Menu